CARDY® Pet:PC-based ECG diagnostic system, version 1.3.0.*
Introduction
Your CARDY® PET intelligent ECG system provides a double way of usage:
- when connecting to a PC, it carries out traditional, laboratory ECG tasks: measurement, analysis, data storing, comparing, printing out of findings
- without PC, it operates as usual stand-alone ECG equipment: measures and analyses ECG
1. Functions of the Cardy system
As a stand-alone equipment:
- Monitoring of 6 limb leads and 2 chest leads
- Storage: raw or processed curves
Connected to PC:
- Monitoring of 6 limb leads and 2 chest leads
- Storage of patients’ data and ECG records on hard drive in compressed form. (On hard drive of 20 Gbyte 1.7 million ECG can be stored)
- Automatic parameter computation and classification
- Visual comparison of ECG measured on different time
- Editing and storage of comment to ECG record by built-in editor
- Printing of findings
- Archiving and recovering data base
- Export and import function for data exchange
- Built-in e-mail sending of ECG record
- Local network supporting
- Easy-to integrate to praxis software (option: uses GDT standard)
- Windows 2000, XP, Vista compatible
1.2 Elements of the system:
- Cardy ECG unit
- PC, infra-cable, software
- Patient cable, electrodes
- Cardy holder (option)
Minimal requirements:
- Pentium CPU, 64MB RAM
- SVGA Monitor, 1 GB HDD
1.3 General requirements
Before starting read this manual carefully. The manufacturer can only be held responsibility for the safety, reliability, and performance of the apparatus, if:
- The electrical installation of the relevant room complies with IEC requirements
- The Cardy is used in accordance with the operating instruction
1.4. Location
Do not keep or operate the apparatus in a wet, moist or dusty environment. Also, avoid exposure to direct sunlight or heat from other sources. Do not allow the unit to come into contact with acidic vapors or liquids, as such contact cause irreparable damage.
The unit should not be placed near X-ray or diathermy units, large transformer or motors. In same cases the switching power supply unit of notebooks and displays can cause 50Hz noise on ECG signal. If it happens, ask assistance from experts.
The Cardy is battery operated, use always 2 pieces of AA type alkaline batteries or compatible accumulators. Take care of polarity when changing the batteries.
Usage of any kind of power supply is dangerous and prohibited.
The patient’s pacemaker or other stimulator can be simultaneously used with Cardy, the do not interfere each other. However the doctor must know the presence of these equipments and he/she should evaluate the ECG in the awareness of this.
The Cardy can not be applied to ECG leads measured directly on the surface of the heart.
1.5. Usage of defibrillator
1.5.1 The input of the Cardy is not protected against defibrillator spike, so during defibrillation the patient cable should be pulled from the Cardy or defibrillation-protected patient cable should be used.
During measurement the patient’s bed, the patient him/herself, and the equipment applied on patient should not be touched as this can lead to injury.
Avoid touching the Cardy, patient or patient cable during measurement, as it can cause noisy ECG.
1.6. A Cardy can not be used during applying high-frequency surgical equipment.
1.7. Before operation
Check if the system is in proper condition with special regard to the connection of the patient cable
Polarization voltage and due to it base line wandering may occur if old and new electrodes are used simultaneously. The same can happen if disposable electrodes are used together with non-disposable electrodes.
1.8. During operation
If the safety of the patient requires, switch off the system or remove the electrodes.
Do not touch the electrodes and the patient. Do not let the patient or the electrodes contact electrically leading units such as metal parts of examining bed, metal stand etc.
During measurement the patient should locate at least 2-3 m from transformer (TV, PC, mobile phone) as these can disturb the measurement.
Special attention should be taken on power supply unit of the notebooks: if the ECG is comprises noise try to pull out the power supply unit from the mains.
1.9. After operation
Detach the electrodes first. The PC must be switched off only if the system will not be used for a longer period of time (some hours)
Clean the non-disposable electrodes after every examination. Use alcohol or other special electrode cleaning liquid for this.
1.10. Maintenance and checking
The system and accessories should be checked regularly by authorized people.
1.11. Specification
Power supply: 3 Volt, battery (2 pieces, AA tip.)
Current consumption: 40mA
Patient protection: floating ground (IEC-601, BF type)
Input range: +/- 5 mV
Sampling: 500Hz, 10 uV/bit (A/D of 10-bite)
Input resistance: 10 MOhm
CMRR: 100 dB
Filters: 100 Hz low pass (hardware, constantly used ),
50 Hz (mains, software, switchable),
30Hz (muscle tremor – myogram - , software, switchable),
Frequency range: 0.05-100Hz
Error of linearity: < 0.5 %
DC tolerance on input: 400 mV
Lead system: 6 limb leads and 2 chest leads (CV5RL, CV6LU)
Installation of the CARDYâ PET PC software
- Insert the setup CD in your CD-ROM driver and wait until the installation program starts automatically. If auto run is disabled on your computer, run “Setup.exe” on the CD. Please type your answers to the questions displayed, click on “Next” ,then “Finish”.
Link to SW downloads:
CardyPet installation:
http://docinfo.hu/Cardy_inst/Cy2Pet_e.zip
b.) Installing the USB IrDA adapter (INFRA port) in Windows XP
- Start Windows XP
- Insert the USB IrDA adapter into a vacant USB port of your system
- Windows should automatically detect the USB IrDA adapter
- Windows will automatically install its built-in driver
This unit connects your CARDY® Pet with your PC. During ECG measurements, this infrared connector is used to send ECG data from CARDY to the PC. The Infrared connector works best if you place CARDY® about 5-10 cms (4-8 inches) in front of the upper infrared window of CARDY®.
Cardy holder is automatically ensures the proper positioning of the infra unit. The cable of infra can be lengthening up to 5 meters with passive cable or more with active USB lengthening cable.
Operation of the Program:
Run “CARDY® Program” by double clicking on its icon. (Generate icon when running the program 1st time from windows Start /Programs menu in the regular way.) The Main Menu will be displayed.
The program functions can be found in the main menu: as text in the top row, and as icons below.
Program menu serves for settings and program exit. The patient menu serves to create new patient file or modify the existing one. The archive or recover the data base and export/import function can be made here, too. Cardy menu handles the Cardy: it reads the ECG files from Cardy in and switches the Cardy off.
The right side text fields correspond to the selected patient. The icons below can be used to erase or modify patient data or to record a new patient.
Most of the program function can be run by hot key (See online help for details)
Press F1 for help.
When you run this program for the first time, start with settings (the Hammer icon)
![]()
The program can work with default settings, so there is no need to change them, but some possibilities serve as comfort for the user.

Settings:
Colors:
Drawings: colors of visualization on display
Doctor’s data: printed on every printout
Language: select English, German or Hungarian
Printer: select the printer you want to use with Cardy program
E-mail: the measurements can be sent by e-mail. The address can be set here.
CARDY® PET PC ECG:
A routine measurement is carried out as follows:
Visualization of 6 limb leads and 2 chest leads on PC display
- Switch on CARDY by pressing the O key. Put the infrared connector in front of the upper IR window of the device. A “ PC USER” message on Cardy LCD will appear. If you want to make a PC measurement, just go to the next step (Run the PC program)
“PC USER” is a special “patient” as no unintentional storage of ECG in Cardy is allowed. If some other user name is seen, ”PC USER” is automatically set by PC, when visualization is started on PC display. If you want to store ECG in Cardy as well, select other user in Cardy” with the ê or é key, and accept it by ü key. Store ECG in Cardy by answering ü for “Save ECG” question after analysis.
- Run the PC program using the “Cardy” icon.
- Check the PC – Cardy connection (“Cardy3 Pet ONLINE” is displayed on the middle bottom line on your PC screen.) If “Cardy offline” is displayed, check Cardy and the infrared connection.
- Record the new patient by clicking the “empty page” icon. Type in the owner’s name and name - this fields are obligatory, the other fields are optional. The Case History is displayed on each ECG records.
Alternatively, select an existing patient from the list. The heart icon turns red (active.)

- Enter the monitoring window by clicking on the Heart icon:
![]()
- Set the leads to be monitored – right-click on any part of the visualization window to select the leads.
- Start monitoring with the Green light icon
![]()
- If the baseline moves, use the block function
- Amplification and shift can be set with the mouse or the arrow keys.
During monitoring, one of the two measurement modes can be selected:
I. Storage of „raw” ECG
II. Storage of raw and analyzed ECG
I. Recording, storage and printing out a max. 60 s long part of an ECG
From 5 to 60 sec. of the monitorized ECG can be stored and printed. At the beginning of the monitoring, a yellow line is displayed that marks the filling up of the „Rough ECG storage (max. 60 sec, 100% is equal with 60 sec.). 5 sec. after the start of monitoring two icons are displayed: a floppy and a printer icon. After this, monitoring can be stopped any time if the ECG curves are to be saved. Both icons become activated as soon as the monitoring stopped. Clicking on the icons the latest 1 minute of the ECG is storable. After 60s the ECG curves are overwritten in the storage. The saved ECG is automatically naPet „5s”, as 5 sec. is the shortest storable fraction. (The program writes the whole stored ECG on the disc, the user can determine the length by ending monitoring. E.g.: if the yellow line shows 50%, the stored ECG is 30 sec. of length.).
The stored ECG can be replayed from the rough ECG storage. The time zone indicator next to the printer icon has four icons:
0s: one click: panning, start/stop; double click: panning start from the
beginning of the acceptable fraction (hotkey: r)
<: moving backward (hotkey: Page Down)
>: moving forward (hotkey: Page Up)
roller: (catching left-click on mouse) positioning
Hotkey for printing: p, hotkey for saving :t
Resume monitoring by clicking on the Green light icon or pressing “s”.
II. Storage of raw and analyzed ECG
Measurement and analysis of a 16 sec ECG, storage of averaged cycles
Start the measurement with the Tick icon.
![]()
Cardy will carry out a 16 second ECG measurement, analysis, and sends the results over to the PC.
The computer displays the ECG curves that can also be printed with the Printer icon, or saved with the PC icon.
After saving the data, you can edit the medical comment with the “Opinion” tab. The opinion is stored along the ECG measurement and appears on the printout as well. You can display the calculated parameters with the “Parameter” tab.
Filters: Both Cardy and PC program have built-in digital filters, which are operating independently. The Cardy filter can be set in Cardy during monitoring by the “O” button. The 1st push of the button switches on the 50Hz mains filter, next push switches on the 30Hz myogram filter (the mains filter is kept on), and the 3rd push switches both filters off. Usage of filter will not prevent the rejection of noisy ECG - as the analysis is carried out from non-filtered signal - , but it improves the quality of the rhythm curve. The setting of filters is kept by Cardy, even if it is switched off. (The ECG stored in Cardy will be filtered, too, when the filters are set on.) When Cardy filter is switched on, all the raw ECG leads and the rhythm curve will be filtered. Note: When Cardy is used with PC, the rhythm curve - ECG of 16s seen after the averaged cycles - is filtered by the Cardy, not by the PC.
The filters in PC program can be activated during monitoring by clicking the proper icon. When clicking the “30Hz” myogram filter icon, it switches on the 50Hz mains filter, too If they are active, the program will store filtered raw ECG in the database.
To display stored data, find the patient (type the first letter of the patient’s name to jump forward in the list), then open the patient data by clicking on the name. Select the date of the recording by double clicking on it. The ECG curves will be displayed.
To compare ECG curves visually, you can select multiple curves of a particular patient (hold down CTRL while clicking on the curves.) You can view the 6 limb and 6 chest leads separately, but always a maximum of 5 curves at the same time. Click “phone-endoscope” icon after selecting the ECG record to be compared.
The complete measurement can be carried out from the keyboard as follows:
Help: F1
Select patient: up/down arrows
New patient: Ctrl+N
Modify patient data Ctrl+M
Save and close patient data Ctrl+S
Discard and close patient data “Esc”
Monitoring window: F2
Start/stop monitoring: s
Block: b
Measurement data gathering during monitoring: “Enter”
Exit monitoring window: “Esc:
Display selected recording: F3
Editing opinion for ECG curve: Ctrl+m
Save and close opinion: Ctrl+s
Zoom±……………………………………………..é or ê
Downloading data from CARDY® Pet Shift+F4
Toggle Cardy type: F5
Printing patient data chart and case history: F6
Settings: F7
Turning Off CARDY®: F8
Functions of the icons, main menu:
![]()
Deletes the selected item (the entire patient record or a measurement)
![]()
Opens data file of an existing patient (modification of patient identity)
![]()
Opens data file of a new patient (creates new patient identity)
![]()
Active, when a patient is selected. Click it to go into the monitoring window
![]()
Visualization of the selected ECG record
![]()
Grab it! Download all ECG’s from a patient selected in Cardy
![]()
Prints patient identity and anamnesis
![]()
Setup
![]()
Switch the Cardy off
![]()
Exit program
Functions of the icons, monitoring:

![]()
Start monitoring (To set the leads to be displayed use right-click of the mouse)
![]()
Stop monitoring
Start ECG recording for analysis
![]()
Block of 1s. If the baseline moves, use the block function
![]()
50Hz mains filter. Click to switch on or off. The status is kept
![]()
30Hz myogram filter. Click to switch on or off. The status is kept
![]()
Time scale setting: 12,5 mm/s, 25 mm/s, 50 mm/s, 100 mm/s,
![]()
Amplification: 5 mm/mV, 10 mm/mV,20 mm/mV,40 mm/mV,

Displays the selected range from stored ECG
Next three icons are active, when the monitoring is stopped by “red lamp” to store or print raw ECG. (Yellow bar on the lower right side of the screen shows the status of the circular buffer: 100% means that the buffer of 1min is full, than the new ECG overwrites the old data.)
![]()
Prints the ECG displayed on display
![]()
Storage of ECG from circular buffer
![]()
Exit monitoring window
Functions of the icons, offline ECG info (displaying of the findings after measurement or stored ECG):

![]()
Print finding
![]()
Zoom +/-
![]()
Time-scale +/-
![]()
Change view of parameter table (Active in “Parameters” menu)
![]()
Save parameters to disk as text or Excel format (Active in “Parameters” menu)
![]()
The menus on the button show different results of findings and offer possibility to edit a comment of the doctor.
CARDY® PET: stand-alone ECG equipment
Operational Manual
CARDY® PET, as stand-alone ECG
You can use your Cardy as stand-alone equipment: the 6 limb leads and 2 chest leads (or 8-lead) ECG signal is displayed on its graphic LCD, and the automata program carries out ECG processing. The measured curves and the averaged ECG cycles can be stored for later PC data transmission.
When buying, the Cardy is ready for measuring and analyses of 8-lead ECG, but further operation modes can be set. (See at “Setup” menu)
A routine measurement is carried out as follows:
Put 6 electrodes on (four limb and 2 chest electrodes) and connect the patient cable to the Cardy. You can reduce no. of used electrodes in setup menu, at Leads item.)
a. Switch on the Cardy: Press the switch on key (O).
à Cardy carries out self test and the patient identity table appears
b. Select patient with ê or é key, and accept patient with ü key. New patient can be added by holding down the ükey and selecting an Icon with ê or é key and accepting it by ü key. (When Cardy is used without PC, do not select “PC USER” as no ECG storage is allowed with this special user name.)
à Main menu of Cardy appears
c. Start visualization of the ECG by answering the question “6 elect’s OK” by ü.
àLead I is visualized.
Change lead by the ê or é key, vary Gain by pressing and holding down the ê or é key
d. Check the ECG signals: select different leads by the ê or é key.
à The ECG leads are displayed one by one
e. Filter: The Cardy program can filter noises with digital filtering: during monitoring use
“O” button to set the “mains” (50Hz) or “myogram” (30Hz lowpass and 50Hz) filter.
Usage of filter will not prevent the rejection of noisy ECG - as the computation of the averaged cycle is carried out from non-filtered signal - , but it improves the quality of the saved ECG file. The setting of filters is kept by Cardy, even if the Cardy is switched off.
f. Start automata program for measurement and analysis: Press ü key.
à ECG of 16 sec is sampled and evaluated.
During measurement all the 6 limb leads and 2 chest leads are displayed one by one automatically.
After analysis the results are displayed.
g. Displaying the results of analysis: press ê key repeatedly to go through the findings
à the averaged cycles of different lead, lead I of 16 sec for checking the rhythm appears
(Use O key to sweep over the rhythm curve.)
h. Storage of measurement: press ü key for “Save ECG record” question while the results are displayed.
àThe actual measurement is stored in Cardy.
i. Exit from measurement: push X in any window showing the result of analysis.
à The main menu appears, where you can switch off by O key.
(Cardy switches off automatically after 3 minutes.)
Displaying the stored ECG
- Switch on the Cardy: Press the switch on key (O).
à Cardy carries out self test and the patient identity table appears
b. Select patient with ê or é key, and accept patient with ü key.
à Main menu of Cardy appears
c. Select “Saved ECG” (directory) menu by pressing ê and ü key
à List of stored ECG appears: type, date and time of records are listed
- Select ECG: by pressing ê or é key
To display the selected ECG, press ü key (or for deleting, press O key.)
à Selected ECG appears. Press ê key repeatedly to go through the findings
(Use O key to sweep over the rhythm curve.)
Saved file erasure
- Switch on the Cardy: press the switch on key (O).
àCardy carries out self test and the patient identity appears
- Select patient with ê or é key, and accept patient with ü key.
àMain menu of Cardy appears
- Select Delete menu by pressing ê and ü key.
àList of deleting mode appears.
- Select Delete Mode by pressing ê or é key and ü key
àAffirmation window for selected deleting mode appears. Confirm it by
pressing ü key. (Note: “Format flash will erasure all the patient records
- See free memory at Delete / Memory status menu item
Check the free memory size before starting to record a rhythm curve of 256s, as this will require 128 blocks of the 354 block total memory capacity.
Memory demands of ECG records in blocks
Leads Operation mode No. of blocks in memory
1 Rhythm, 32 sec: 16
Rhythm+ ,256 sec: 128
6 4sP: Raw ECG of 4s and rhythm curve: 12
8 Pet: averages and rhythm curve of 16s 16
Pet+: averages, rhythm curve, raw ECG (2s) 20
Raw ECG of 4s and rhythm curve: 16
Maximal no. of patient identity: 10 (“PC USER” + 9 selectable)
Maximal no. of stored records: up to 48 (in basic model, in “4sP” mode)
Restarting Cardy (RESET) Complex electronic devices in Cardy are sensitive to external noise such as static electricity, or electromagnetic waves close to electronic devices. If for any of these reasons Cardy should “freeze” – meaning that the display would not change after 30 seconds, you can reset (reboot) Cardy as follows:
- Press and hold the ü key.
- Press and release the switch on key (O).
- Release the ü key.
Setup menu
The Mode, the Leads, The Rhythm and the Icon menu items are specific: they refer to the selected patient. All other menu items are general. For “PC USER” the Mode and Leads are fixed for 8-lead analysis, it can not be changed: use other patient for different mode/lead.
Set ECG measurement type by selecting Mode menu withê or é key, and ü key.
àModes appear:
Pet: measurement of the leads set in Lead menu (8-lead ECG) and storage of
averages and rhythm curve (16 sec from one lead)
Pet +: measurement of the leads set in Lead menu (8-lead ECG) and storage of averages, rough ECG of 2 second from measured leads and rhythm curve (16 sec from lead I)
Rhythm 32s: measurement and storage of one lead of 32 sec
Rhythm+ 256s: measurement and storage of one lead of 256 sec
Raw ECG, 4s: measurement and storage of 4s from all leads set in Lead menu
Select the required Mode by ê or é key, and accept it with ükey.
Set no. of leads to be measured by ükey at “Leads” menu item.
Rhythm: the lead for rhythm curve with analysis mode can be I, II, CV5RL or CV6LU.
Set the sound on/off by selecting the “Sound” menu item and pressing the ükey.
Set date and time by selecting the Date/Time menu item and using ê or é and ü. Language selection is available in Language menu item.
Login icon for selected patient can be chosen at Icon menu by ê or é and ü key.
Add patient and Remove patient menu items can be activated from patient identity window (the 1st window after switching in) by pressing and holding down the ü key. The former creates a new patient file for coming measurements, while the later erasures the selected patient and all of its ECG records.
Data transfer of stored ECG from Cardy bedside ECG to PC
Select the patient the ECG records of whom are to be transferred from the Cardy to PC by ê or é and ü key.
àCardy displays the selected patient’s menu window
Run Cardy PC program and put Cardy in front of the infraport.
Select patient from database or create a new the stored data in Cardy refers to.
Use ![]()
icon to transfer ECG record from Cardy to PC.
CARDY® Pet:Frequently asked questions
Program installation and usage
How many installations can be made with a software: as many, as you wish.
What happens, if reinstalling the Cardy program: will it overwrite the data? No, all the measurements will be preserved.
Error message during installation: “16- bit system does not support application of 32-bit”
Reason: No autoexec.nt file was found in directory: Windows\system32
What to do: Copy autoexec.nt file from Windows\repair directory to Windows\system32 directory
The windows of the program has not proper size (some bottoms /text are missing)
Reason: No 96dpi is set on display
What to do: Set Normal (96DPI) in Display/ Settings/Special/ General menu
The visualized chest leads on PC display are not “ECG like”:
Reason: the Cardy operation mode is set to rhythm or rhythm+ where the Cardy doesn’t transmit chest leads to the PC
What to do: set “P”, “P+” (8 –lead ECG)
The “Cardy/Download new firmware” menu does not offer web-site address.
Reason: this menu is a tool to update the Cardy firmware, not the PC software.
What to do, if you wonder, whether a new firmware exists: ask your dealer.
The Cardy can not be switched on:
Reason1: No battery in or no energy in batteries
Reason2: no proper connection at batteries, check it
Reason3: the equipment has not been used for a long period of time, and the contrast is forgotten because of the lack of battery
What to do: After changing the battery, set the contrast: hold down the O key and down arrow key simultaneously (brightening) or up arrow key simultaneously (darkening)
Checking FS, Error 0x0281 error message on Cardy after switching on, or
Error message “Temp. write protected” before ECG storage
Reason: no flash formatting was made after firmware update
What to do: Format flash in Delete menu, Format Flash:
Be careful: all the data will be lost in
Will the data be kept in Cardy during battery change? Yes
Measurement error during analysis
Reason: During measurement Cardy checks the placement of the electrodes and the technical quality of the ECG. If the electrodes are not connecting well during the measurement, or the ECG is “noisy”, an error message will appear after the ECG sampling:
„High noise CODE: F8” (Most frequent error code is F8, but it can be any error between FF and F0.) Check the placement of the electrodes, check if all the pins are in place, make sure that you are not too near to an electronic device, such as a TV,PC, or especially a power supply of a note-book.(If you use note-book, pull out its power cord from mains during measurement, if the ECG signal is noisy). If everything seems to be in order, press “YES” key, the patient has to relax, and the measurement can be repeated. Note: the filter is used to get noise-free signal on LCD, but the analysis is carried out from unfiltered ECG signal, so you should find the reason of noise and eliminate it.
Noisy ECG, when the ECG record is made by doctor/second person (called: “assistant”)
Reason: the assistant disturbs the measurement by his own electromagnetic field.
What to do: the assistant should be located at a minimal distance of 1 m from the patient during measurement. The assistant should not touch the patient or patient cable during measurement
Communication error during monitoring or downloading
when the processor time of the PC is taken away by some other program, the infra communication can have malfunction.
- do not allow internet communication during Cardy communication (monitoring, or on-line ECG measurement). Windows upgrade is very "selfish": takes the CPU time away for a long time!
- switch off antivirus program (no need for it, if internet is disabled)
- the network communication may disturb Cardy communication
Where to find Cardy install program:
English version: http://docinfo.hu/Cardy_inst/Cy2Pet_e.zip
CARDY® Pet: Meaning of messages and user operation
Select menu item by ê or é key, and enter/activate menu item by ü key. Escape from menu by X (no) key.
Cardy/g messages Meaning, what to do
(After switching on)
CONTACT SERVICE The self test resulted in a software or hardware failure, Contact service
Real-time clock error The clock has to be set:
Pushing down ê,é : sets one digit,
Holding the ê,é : increases/decreases continuously
ü key: accepts, jumps to next,
X key: steps back
Date and time set After setting the date/time, Cardy acknowledges
Please upload Petia No sound file found.
file! Contact service
Select patient After switching on, select the PC USER/PATIENT 1 patient to be examined.
PC USER: Cardy Pet used with PC
The ECG can be stored only in PC
PATIENT 1 PATIENT 1…9:
The ECG can be stored in both Cardy and PC
MA+ N /Y /R Operation mode ECG’ s / yellow /red alarms
ê é ü Selects Push ê or é to select patient and then accept it by ü
Hold ü Menu To create/remove a patient: push and hold ü
Patient add/removal table:
Select the required function by
ê or é and accept it by ü
files deleted After successful delete this message is seen
Patient removed The selected patient is removed from Cardy
Maximum number of Eight different patients may exist simultaneously
patients reached
Last patient cannot be Cardy keeps at least
removed one patient
(Main menu) After patient selection, the main menu appears
“N” elect's OK? MA Answer ü to start monitoring.
N: No. of electrodes to be used,
MA: operation mode: P (8-lead ECG),
Saved ECG's Select this menu to display stored ECG
Delete Selected ECG (‘s) can be deleted in this menu
Setup Operation mode, No. of leads, lead selection for rhythm , sound on/off, date/time, language, user icon can be set
(monitoring)
Analysis is rejected because of noise. By answering ü,
raw ECG of 4s can be stored.
(delete menu) Delete menu offers different possibilities
No change This menu deletes all ECG’s where no changes were found
All but ref This menu deletes all ECG’s
All This menu deletes all ECG’s
Format flash This menu deletes all ECG’s of all patients
Every delete function asks for affirmation
files deleted After successful delete this message is seen
(setup)
(Contrast) This message cannot be seen, but any time the contrast can be set: hold down the O key and push ê or é key
Mode To set the required operation mode, push ü. The modes are:
Storage of averages and 16 sec of one lead
Storage of averages, rhzthm curve and raw ECG of 2 sec of all measured leads are alsostored
32 sec of lead I is stored to check rhythm
256 sec of lead I is stored to check rhythm
ECG of 4 sec is stored (no analysis)
Leads Select 8-lead (Pet) or -6 lead (raw ECG) by ü
Rhythm One lead of 16 sec is stored in Pet and Pet+ mode to check rhythm. Select this lead:
enter the lead table by pushing the ü key.
Lead I,II, CV5RL or CV6LU can be selected by ê or é key
and ü
Sound Use ü to set On/Off/Keys and exit with “X”
Date/time Select this menu item by ü to set date/time
Pushing down ê,é : sets one digit,
Holding the ê,é: continuously increases/decreases,
ü key: accepts and jumps to next,
Language Selects one of the 5 languages
Select with ê or é key and accept it by ü key
Icon Select icon what you like to identify patient
Select new login icon! You can modify login icon any time.
PC user cannot be No way in CardyPet to cancel the “PC USER”
removed
Useful Tips
Switching off: Cardy switches off automatically after 3 minutes of inactivity , but you can manually switch off Cardy by holding pressed the O key for 3 sec’s. (Auto-switch is not active before ECG storage.)
Replacing the battery: A standard alkali battery has an approx. 60 hours (approx. 720 measurement) lifetime. A 1200 mAh rechargeable battery can power Cardy for about 30 hours (360 measurements). An icon shows the battery voltage.
Contrast setting: hold down the O key and ê key simultaneously (brightening) or ékey simultaneously (darkening)
Communication error during monitoring or downloading
when the processor time of the PC is taken away by some other program, the infra communication can have malfunction.
- do not allow internet communication during Cardy communication (monitoring, or on-line ECG
measurement). Windows upgrade is very "selfish": takes the CPU time away for a long time!
- switch off antivirus program (no need for it, if internet is disabled)
- the network communication may disturb Cardy communication
Electrode positions
Red : right anterior leg
Yellow: left anterior leg
Green: left posterior leg
Black: right posterior leg
White: -CV5RL, right chest lead
Blue -CV6LU, left chest lead
